• we.webp

QE Quantum Efficiency Tester - TAB-401


The TAB-401 QE Quantum Efficiency Tester is capable of measuring the quantum efficiency (photoelectric conversion efficiency/spectral response characteristics) of various photovoltaic materials such as solar cells. It is a complete turnkey testing solution that enables automated measurements, complying with major international photovoltaic testing standards including ASTM (American Society for Testing and Materials) and IEC standards. Suitable for both standard laboratory research and quality control measurements in industrial production.


Keyword:

QE Quantum Efficiency Tester - TAB-401

PRODUCT INTRODUCTION

The TAB-401 QE Quantum Efficiency Tester is capable of measuring the quantum efficiency (photoelectric conversion efficiency/spectral response characteristics) of various photovoltaic materials such as solar cells. It is a complete turnkey testing solution that enables automated measurements, complying with major international photovoltaic testing standards including ASTM (American Society for Testing and Materials) and IEC standards. Suitable for both standard laboratory research and quality control measurements in industrial production.

Testable Cell Types: Includes various solar cells and detectors, such as silicon-based solar cells, thin-film solar cells, dye-sensitized solar cells, new perovskite solar cells, multi-junction solar cells, etc.

 

PRODUCT FEATURES

●  Test Light Source: Irradiates from top to bottom for convenient sample placement and measurement.

●  Light Emission Direction: Also configurable for horizontal irradiation to accommodate various cell types.

●  Effective Measurement Module Size: Adjustable.

●  Compatible with Nanowire Solar Cells: Capable of testing nanowire solar cells.

●  Measurement Area Range: 0.01mm² ~ 200×200mm² (or larger area optional).

●  Expandable Function: Can be expanded to measure photovoltaic modules with power not less than 200W.

●  Typical Measurement Spot Size: 1×1mm (minimum up to less than 0.01mm²).

●  Loading Method: Offline manual loading.

 

PARAMETER SPECIFICATIONS

Equipment NameQE Test System - TAB-401
Spectral Range200-1100nm; Expandable to 1800nm
Typical System Test Wavelength Range300nm-1100nm; (Expandable to 1800nm or 200-2500nm as an option)
System Test Result Repeatability≤ 0.15% rms (300–1100nm), ≤ 0.1% rms (1100–1800nm)
Stray Light<10⁻⁵
Wavelength Accuracy<0.1nm
Standard Test Sample Stage(Optional temperature-controlled vacuum adsorption test sample stage: Size 200*200mm; Temperature control range: -15-60℃; Temperature control accuracy: 0.1℃)
IPCE Test System SoftwareFunctions include testing spectral response, reflectance, transmittance, internal quantum efficiency (IQE), external quantum efficiency (EQE), IPCE (Incident Photon-to-Electron Conversion Efficiency), and current density of samples;Draws curves and saves test data (supports file formats such as Excel and TXT);Computer control: Automatic scanning, motorized slits, grating replacement, filter wheel control, data acquisition;Data processing: Spectral calibration, curve plotting, data analysis, data storage, report printing.

 

●  Test Light Source: Irradiates from top to bottom for convenient sample placement and measurement.

●  Light Emission Direction: Also configurable for horizontal irradiation to accommodate various cell types.

●  Effective Measurement Module Size: Adjustable.

●  Compatible with Nanowire Solar Cells: Capable of testing nanowire solar cells.

●  Measurement Area Range: 0.01mm² ~ 200×200mm² (or larger area optional).

●  Expandable Function: Can be expanded to measure photovoltaic modules with power not less than 200W.

●  Typical Measurement Spot Size: 1×1mm (minimum up to less than 0.01mm²).

●  Loading Method: Offline manual loading.

Equipment Name

QE Test System - TAB-401

Spectral Range

200-1100nm; Expandable to 1800nm

Typical System Test Wavelength Range

300nm-1100nm; (Expandable to 1800nm or 200-2500nm as an option)

System Test Result Repeatability

≤ 0.15% rms (300–1100nm), ≤ 0.1% rms (1100–1800nm)

Stray Light

<10⁻⁵

Wavelength Accuracy

<0.1nm

Standard Test Sample Stage

(Optional temperature-controlled vacuum adsorption test sample stage: Size 200*200mm; Temperature control range: -15-60℃; Temperature control accuracy: 0.1℃)

IPCE Test System Software

Functions include testing spectral response, reflectance, transmittance, internal quantum efficiency (IQE), external quantum efficiency (EQE), IPCE (Incident Photon-to-Electron Conversion Efficiency), and current density of samples;Draws curves and saves test data (supports file formats such as Excel and TXT);Computer control: Automatic scanning, motorized slits, grating replacement, filter wheel control, data acquisition;Data processing: Spectral calibration, curve plotting, data analysis, data storage, report printing.

Related Products


Submit
%{tishi_zhanwei}%

Leave Message

If you have already experienced our product, please let us know your true feelings. Your satisfaction is our driving force for progress, while your suggestions are our valuable asset for continuous improvement.

Contact Us

WhatsApp

中企跨境-全域组件 制作前进入CSS配置样式

在线客服添加返回顶部

右侧在线客服样式 1,2,3 1

图片alt标题设置: Wuhan Aijiang Technology Co., Ltd.

表单验证提示文本: Content cannot be empty!

循环体没有内容时: Sorry,no matching items were found.

CSS / JS 文件放置地

Welcome to leave an online message, we will contact you promptly

%{tishi_zhanwei}%
gtag('config', 'AW-17914789977');